• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS ANTI-HBE; ANTI-HBE TEST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS ANTI-HBE; ANTI-HBE TEST Back to Search Results
Catalog Number 07026838190
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The analyzer serial number is (b)(6).The investigation is ongoing.
 
Event or Problem Description
There was an allegation of false positive elecsys anti-hbe results for 1 patient sample on a cobas e 801 analytical unit.The initial result was 0.637 coi (reactive).The repeated result was 0.648 coi (reactive).Calibration was performed and the sample was retested.The results after calibration were 0.743 coi (reactive) and 0.718 coi (reactive).The results did not fit the clinical picture of the patient.The sample was tested using another method (abbott alinity) and the result was 2.06 (non-reactive).
 
Additional Manufacturer Narrative
A sample from the patient was requested for investigation, but the customer declined to send one.The investigation could not identify a product problem.The reagent performs within specifications.Single false positive results can occur and are covered by the claimed specificity for anti-hbe assay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-HBE
Common Device Name
ANTI-HBE TEST
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18873768
Report Number1823260-2024-00720
Device Sequence Number11532115
Product Code LOM
UDI-Device Identifier04015630940134
UDI-Public04015630940134
Combination Product (Y/N)Y
Initial Reporter CountryTH
PMA/510(K) Number
P190005/S002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue Number07026838190
Device Lot Number68937201
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/15/2024
Supplement Date Received by Manufacturer03/13/2024
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date03/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-