• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON TRAY SPN WHIT25G3.5 L/B-D/E BLUE DRAPE; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON TRAY SPN WHIT25G3.5 L/B-D/E BLUE DRAPE; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 405672
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4): initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
Material #405672.Lot #unknown.It was reported by customer that spinal anesthesia tray did not properly numb the patient on three separate occasions from the same lot.Additional information: could you please provide a further description of the issue? the anesthesiologist was able to administer the spinal anesthesia.The patient responded as expected to the anesthesia with decreased bp, however, the spinal did not set up to numb her sensory for the procedure.She was unable to move her legs, but could still feel everything.This happened with at least 5 of the spinal trays with the same lot #.2.Can you please provide the lot number associated with reported issue? (b)(6), can you provide this information? 3.Can you please provide a exact date of event? 2/9/2024.What was the impact to the patient? was there any injury? if so, please describe and include any medical treatment needed as a result? patient required general anesthesia after the spinal anesthesia didn¿t provide adequate coverage.
 
Additional Manufacturer Narrative
One empty tray and two assembled trays were returned.No defects or issues observed.A device history review was performed for lot 0001525982 no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Retained samples from the same lot were evaluated, no issues or defects observed.The sterilization release record was reviewed for the finished good trays, the requirements for it's release were met.Based on the investigation results, no manufacturing related defects could be identified and therefore, a cause for the reported incident could not be determined.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures.Manufacturing personnel have been notified of this incident to increase awareness.
 
Event or Problem Description
Additional information: 5.How was the issue resolved? all spinal trays with the affected lot number were pulled from our ors and given to our facility materials management team 6.Is there a photo or sample available showing the reported issue? if yes, are you able to provide the address of the facility for us to ship the return label? steve may be able to provide this image.Lot numbers are: 0001525982 plus one with only the front.I have 3 samples that can be sent back.Our address is: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAY SPN WHIT25G3.5 L/B-D/E BLUE DRAPE
Common Device Name
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18874503
Report Number1625685-2024-00026
Device Sequence Number9069094
Product Code CAZ
UDI-Device Identifier00382904056728
UDI-Public(01)00382904056728
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue Number405672
Device Lot Number0001525982
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2024
Supplement Date Received by Manufacturer04/23/2024
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date04/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-