MAUDE Adverse Event Report: COOK INC THAL-QUICK CHEST TUBE SET; GBX CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC THAL-QUICK CHEST TUBE SET; GBX CATHETER, IRRIGATION Back to Search Results
Model Number N/A
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the wire guide from three thal-quick chest tube sets unraveled.At this time, no adverse effects or additional procedures for the patient were reported due to this occurrence.Additional information regarding the event and patient outcome has been requested but is currently unavailable.
 
Additional Manufacturer Narrative
G4 ¿ pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Additional information was provided on (b)(6) 2024 detailing that the three incidents of the wire guide fraying all occurred on the same patient on the same day.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: b5 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Two device labels and one wire guide were returned to cook.The two labels indicate that the three reported complaint wire guides came from different device lots.This report will capture the wire guide from the initial lot reported.The two other devices will be capture in reports with patient identifiers (b)(6).
 
Additional Manufacturer Narrative
Additional information: b5, d9.This report includes information known at this time.A follow-up report will be submitted should additional, relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation.It was reported that the wire guide from three separate thal-quick chest tube sets unraveled during the same procedure.This report captures one of the sets, while patient identifiers (b)(6) capture the other two sets.No adverse effects have been reported.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One used wire guide was received.The wire guide started to unravel at the distal solder joint.Both weld balls were present and intact.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the dhr for the reported lot and related wire guide subassembly lots, recording one relevant nonconformance for coil tension in which one device was scrapped prior to further processing.A complaint history search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The product ifu, [c_t_thal_rev11] ¿thal-quick chest tube sets and trays,¿ provides the following information to the user related to the reported failure mode: ¿instructions for use -when the appropriate drainage site has been identified, advance the soft ¿j¿ end of the wire guide through the needle and into the pleural space.Note: the wire guide should pass through the needle and advance into the pleural space without resistance.Note: the skin level marker on the distal end of the wire guide can be used as an insertion depth marker on patients with normal anatomy.Note: due to anatomical variations in chest wall thickness, insertion depth of introducer needle, wire guide and dilators will vary from patient to patient.-remove the needle, leaving the wire guide in place.-liberally inject additional local anesthetic into the intercostal muscles surrounding the wire guide.-while maintaining wire guide position, dilate the tract and opening into the pleural space by advancing, in sequence small to large, the supplied dilators over the wire guide.Introduction into the pleural space is facilitated by rotating and advancing the dilators in line with the wire guide to prevent its kinking.Note: it is important to have enough length of wire guide exiting the access site to facilitate controlled introduction of the dilators.The dilators only have to enter the pleural space to their full diameter and should never be advanced beyond the distal end of the wire guide.-with the wire guide still positioned within the pleural space, advance the chest tube inserter/chest tube assembly over the wire guide and into the pleural space.Note: if resistance is encountered during chest tube assembly insertion, determine the cause of resistance and take necessary action to relieve resistance before proceeding.-note: it is important to advance the chest tube assembly into the pleural space in the same line as the wire guide.This will make introduction easier and avoid kinking of the wire guide.Note: the distal end of the chest tube inserter should not be advanced beyond the distal end of the wire guide.-remove the wire guide and chest tube inserter leaving the chest tube in place.¿ based on the information provided, inspection of the returned device, and the results of the investigation, cook has concluded that component failure unrelated to manufacturing or design deficiencies contributed to this incident.It is possible that the wire guide was manipulated through the needle, resulting in the wire guide damage.However, this cannot be confirmed without additional information.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THAL-QUICK CHEST TUBE SET
Common Device Name
GBX CATHETER, IRRIGATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18875244
Report Number1820334-2024-00311
Device Sequence Number9050678
Product Code GBX
UDI-Device Identifier00827002048911
UDI-Public(01)00827002048911(17)270122(10)15858077
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberC-TQTS-1600
Device Lot Number15858077
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/05/2024
Supplement Date Received by Manufacturer04/04/2024
05/23/2024
06/04/2024
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date04/08/2024
05/24/2024
06/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-