| Catalog Number |
DWD026 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the tightening of a dwd026 locking screw for the reverse ii glenoid system, the screw's head broke cleanly.Despite being sufficiently tightened with no additional effort required and the incident was reported.The surgery was completed without further complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed, since the device was returned for evaluation, and matches the alleged failure.The device inspection revealed the following: -the multidirectional screw is broken at the first threads of the screw head.-the hexagonal footprint is well deformed at the screw head.-a visual microscopic inspection showed that the device rupture corresponded to a fracture caused by a torsional overload, indicating that the device was subjected to high force during use.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by application of a high torsional force that contributed to the reported event.If any further information is provided, the complaint report will be updated.
|
| |
|
Event or Problem Description
|
|
During the tightening of a (b)(6) locking screw for the reverse ii glenoid system, the screw's head broke cleanly.Despite being sufficiently tightened with no additional effort required and the incident was reported.The surgery was completed without further complications.
|
| |
|