| Model Number |
4.75MM BC KNOTLESS SWIVELOCK |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 2/21/2024, it was reported by a sales representative via email that an ar-2324kbcc biocomposite knotless swivelock mechanism locked up and would not tension.The surgeon ended up cutting the repair sutures and removing the anchor.The case was completed using another ar-2324kbcc biocomposite knotless swivelock.This was discovered during an rcr procedure on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misuse/mishandling due to damage to the device.
|
| |
|