Model Number 4K LAPAROSCOPE,0¿, 10X330MM, NIR
Medical Device Problem Code Fogging (1253)
Health Effect - Clinical Code Liver Laceration(s) (1955)
Date of Event 12/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a laparoscopic cholecystectomy excessive fogging was noticed on the scope.The surgeon decided to continue to the surgery with the device.During the surgery however due to obstructed view the surgeon caused a laceration to the liver due to the fogging.The patient was kept overnight to be monitored but no further treatment was reported.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 since the device was not received, an evaluation on the product could not be performed and the reported event cannot be verified.Obstructed views are most likely resulting from damages at the distal end and fiber optics caused by improper device handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K LAPAROSCOPE,0¿, 10X330MM, NIR
Common Device Name
LAPAROSCOPE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key18877363
Report Number1220246-2024-01411
Device Sequence Number12731792
Product Code GCJ
UDI-Device Identifier00888867161115
UDI-Public00888867161115
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4K LAPAROSCOPE,0¿, 10X330MM, NIR
Device Catalogue NumberAR-3351-1001
Device Lot Number12235133
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2023
Supplement Date Received by Manufacturer12/13/2024
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date01/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient SexUnknown
-
-