| Model Number |
AVEA |
| Medical Device Problem Code |
Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that fraction of inspired oxygen set to 40% reading 60% on the avea ventilator.The customer reported there was no patient involvement associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: vyaire medical field service technician went onsite to evaluate the device.All work performed in accordance with avea service manual revision g.Technician performed est and performance check all passed.Ventilator is operational and meets manufacturer specifications.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|