• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; STERILE DISTAL END CAP WITH ELEVATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; STERILE DISTAL END CAP WITH ELEVATOR Back to Search Results
Model Number OE-A63
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
F10 continued: international medical device regulators forum (imdrf) adverse event reporting health effect clinical code: 3165 device embedded in tissue or plaque health effect impact code: 4641 unexpected medical intervention medical device problem code: 2907 detachment of device or device component component code: 424 cap _________ customer did not record any of the lot numbers in any of the patient charts when this occurred.At this time they do not have traceability on the distal caps on which lot number was used.Per the account, they only have two lot numbers of sterile distal end cap model #: oe-a63 on site.Lot #: 0011112 & 0021122.Therefore, the d4 lot number is listed as #: 0011112 and 0021122, two lot numbers.Pentax medical america performed a good faith effort to gather additional information regarding this event and provided an email, but has not received a response at this time.Investigation is in-process.If additional information becomes available, a supplemental report will be filed with the new information.Importer mdr 2518897-2024-00019, dec distal cap model oe-a63, lot number 0011112 or 0021122 importer mdr 2518897-2024-00020, duodenoscope model ed34-i10t2 serial number unknown.
 
Event or Problem Description
Pentax medical was made aware of a complaint on 22-feb-2024 that occurred during treatment in the united states involving a pentax medical sterile distal end cap(dec) model oe-a63, lot number 0011112 or 0021122.The sterile single use distal cap was used with pentax medical video duodenoscope, model ed34-i10t2, unknown serial number.The customer reported that after a procedure the oe-a63 sterile distal end cap came off inside of the patient's stomach.They had to stop the procedure, in order to retrieve the oe-a63 sterile distal end cap.This was not reported previously by the user facility's endoscopy technician supervisor.The exact date is unknown at this time, but customer stated it was approximately 22 months ago (april 2022).The customer also did not record any of the lot numbers in any of the patient charts when this occurred.At this time they do not have traceability on the decs lot numbers used.Per the account, they only have two lot numbers of sterile distal end cap model #: oe-a63 on site.Lot #: 0011112 & 0021122.B3: the exact date of occurence is unknown at this time, but customer stated it was approximately 22 month ago (april 2022).Therefore, the b3 date of event was listed as 01-jan-2022.This event meets the requirements for fda reportability; however, submission of this report does not constitute an admission that medical personnel, user facility, importer, manufacturer or product caused or contributed to the event.
 
Event or Problem Description
Refer to h11.
 
Additional Manufacturer Narrative
Correction information: b4: date of this report; b5.Refer to h11; f7: follow up #01; f11: updated dates; f13: updated dates.Additional information: d4: unique identifier (udi) corrected; h11: evaluation summary.Evaluation summary: pentax medical america performed a good faith effort to gather additional information regarding this event and provided an email, but had not received a response.It was determined that the distal end cap (dec) was not properly attached by the user, and that the dec most likely fell off in the stomach due to the load applied to the tip when it was inserted into the patient's duodenum.Pentax medical has not received any further information for this event and therefore, considers this medwatch report closed.Importer mdr 2518897-2024-00019, dec distal cap model oe-a63, lot number 0011112 or 0021122 importer mdr 2518897-2024-00020, duodenoscope model ed34-i10t2, serial number (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
STERILE DISTAL END CAP WITH ELEVATOR
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
MDR Report Key18879712
Report Number2518897-2024-00019
Device Sequence Number9050808
Product Code FDT
UDI-Device Identifier04961333232994
UDI-Public04961333232994
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
Report Date (Section F) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberOE-A63
Device Lot Number0011112 OR 0021122
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/22/2024
Event Location Hospital
Date Report to Manufacturer05/20/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date05/20/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA05/20/2024
-
-