| Catalog Number |
1012475 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when removing the ht winn guide wire from its packaging it was observed that the distal part of the guide wire tip had no coating and the coils were exposed.Another guide wire was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection was performed on the returned device.The reported peeled / delaminated could not be confirmed.A review of the corrective and preventive actions (capa) database was performed and revealed no related complaint assessment nonconformities.The electronic lot history record (elhr) revealed no associated manufacturing nonconformities issued to the reported lot.Additionally, a query of the complaint handling database for the reported lot revealed no indication of a lot specific issue.The investigation determined the reported peeled / delaminated appears to be related to the operational context of the procedure.In this case, it was reported that the polymer was missing from the distal tip.Analysis of the returned wire noted no polymer separated.It should be noted that the winn guide wire specifications state that the tip of the wire has exposed coils for 4-7 mm.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|