• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3507170MC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 08/21/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 39-year-old caucasian female patient underwent primary breast augmentation with a 170cc mentor memorygel breast implant and experienced breast implant rupture on her right side postoperatively.The patient underwent replacement surgery with allergan devices on (b)(6) 2024.
 
Additional Manufacturer Narrative
Photo evaluation summary: upon visual evaluation of the image provided in the complaint, the implant was observed ruptured in two pieces.As the product involved in this complaint was not received, the device couldn´t be analyzed according to our procedures.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Reason for device explant and/or reoperation: right rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2024, the mentor failure analysis lab received the device for evaluation.On (b)(6) 2024, device evaluation was completed as follows: mentor conducted a visual inspection, microscopic examination, and thickness measurement of the returned device.Visual analysis of the returned sample revealed that the sm mod classic gel, 170cc breast implant was found to be ruptured, received in two (2) parts.Microscopic examination was performed, and the cause of the rupture could not be identified.Therefore, a thickness measurement was conducted on the shell at the tear, and the thickness was within manufacturing specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Although no conclusion could be reach on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Some breast ptosis is a normal component of the mature breast.Ptosis becomes undesirable when the breast parenchyma predominates below the areola, droops considerably below the inframammary fold and the nipple points downward.Augmentation alone, without consideration of the ptosis can produce a less than desirable cosmetic result known as a "rock in sock" deformity or an increased ptotic appearance.Ptosis is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key18880408
Report Number1645337-2024-02978
Device Sequence Number21094937
Product Code FTR
UDI-Device Identifier00081317004080
UDI-Public00081317004080
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/13/2020
Device Catalogue Number3507170MC
Device Lot Number6934546
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/21/2024
Supplement Date Received by Manufacturer03/12/2024
Initial Report FDA Received Date03/11/2024
Supplement Report FDA Received Date03/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/13/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-