| Catalog Number |
3507425MC |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
02/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation revision procedure with a mentor memorygel breast implant 425cc gel breast prostheses that both ruptured post implantation.Bilateral rupture was diagnosed via ultrasound.As a result, the patient is scheduled to undergo bilateral explantation on (b)(6) 2024.This medwatch report is for the patient¿s left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture.D6b explantation month, day, and year: (b)(6) 2024.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 10-jun-2024, mentor received additional information about the event.The patient decided to postpone the explantation surgery scheduled for (b)(6) 2024.An updated explantation date was not provided.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 18-sep-2024, mentor received additional information about the event: the patient underwent bilateral explantation and replacement with a mentor memorygel breast implant 425cc gel breast prosthesis and a mentor memorygel breast implant 405cc gel breast prosthesis on (b)(6) 2024.During explant surgery, both removed breast prostheses were noted to be intact.The removed left breast prosthesis had two folds on it.Block b1 ¿is product problem¿ no longer applies to this report nor does h6 medical device problem code material rupture (a0412).Reason for device explant and/or reoperation: ultrasound diagnosed breast prosthesis rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 02-oct-2024, the mentor failure analysis lab received the device for evaluation.On 04-oct-2024, mentor completed the investigation on the suspect medical device.Device evaluation summary: according with the information received, the breast prosthesis was suspected to have ruptured.During explant surgery, it was discovered to be intact.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|