| Model Number |
HUMERAL LINER, 40MM, +2.5 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
02/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
D10: concomitants: a798874 - 300-01-11 - equinoxe, humeral stem primary, press fit 11mm 7232012 -320-10-05 - equinoxe reverse tray adapter plate tray +5 a566398 - 320-20-00 - eq reverse torque defining screw kit additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
It was reported that a 72 yo female patient, initial right shoulder implanted on (b)(6) 2023, underwent a revision procedure on (b)(6) 2024, approximately 2 months post the initial procedure.The patient was revised due to infection.The surgeon performed a washout and replaced the liner.There were no surgical delays or device breakages during the procedure.The patient was last known to be in stable condition following the event.The rep was unable to obtain any device images or x-rays.The explanted device is not available for return as it was disposed of by the hospital.No further information.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the infection and subsequent revision cannot be conclusively determined.The following sections have corrected information: h6: component code.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h3, h6, h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.A review of complaint history determined that no further action is required as no trends were identified.Neither corrections nor corrective actions are required because the possible risks related to this event are within the level assessed in the current risk management file and "infection¿ is listed in the product labeling [ifu 700-096-060 rev r].This reported failure does not appear to be a systemic issue.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the infection and subsequent revision cannot be conclusively determined should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h6 medical device problem code, types of investigation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h11.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.A review of complaint history determined that no further action is required as no trends were identified.Neither corrections nor corrective actions are required because the possible risks related to this event are within the level assessed in the current risk management file and "infection¿ is listed in the product labeling [ifu 700-096-060 rev r].This reported failure does not appear to be a systemic issue.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the infection and subsequent revision cannot be conclusively determined.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|