This report is being supplemented to provide additional information received through follow-up, correction to the initial medwatch for information inadvertently left out (a2, e2, e3, d8), correction to incorrect date entered in d9, and to provide updates (b1, b2, b5, h1, and h6).
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additionally, to provide an update to fields h3 and h4.The device was evaluated by olympus, and the breakage of the probe was not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and past investigation results, it is likely the probe was broken occurred due to contact with a surgical instrument or the non-insulated area of grasping section.The detailed mechanisms are shown below.1.During output in seal & cut mode, the probe came in contact with hard tissue, metal or a surgical instrument.Consequently, a scratch was made on the probe.2.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.3.The probe broke when added load.1.The distal end of the tissue pad was worn away because ¿seal & cut¿ output was activated while grasping nothing (including the case the user kept activating after cutting tissue).2.The tissue pad was worn away, causing the non-insulated of the grasping section to touch the probe.3.¿seal & cut¿ output was activated under this situation, then the scratches indicating that the probe and grasping section were in contact with each other were made.4.The probe received an output load in seal & cut mode or received a load when grasping tissue.As a result, the probe cracked from the scratch.5.The probe broke when added load.However, a specific root cause of the reported event could not be determined.The event can be detected/prevented by following the instructions for use which state: "treatment of hard or thick tissue in the seal & cut mode may result in failure to cut/coagulate because of the overload imposed on the probe tip." "when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation." "when a body fluid or tissue is present on the grasping section, probe tip or shaft surface during the procedure, immediately remove it by immersing the grasping section and probe tip in saline or wiping with sterile gauze.Otherwise, the performance may be degraded.To avoid the load increasing around the distal end of the grasping section in order not to become hard to open or close, try to keep a clean grasping section during treatment by users.Otherwise vibration failure or other issues may occur." "when treating a blood vessel and/or tissue, squeeze the control handle firmly until the control handle touches the grip handle to stop.Otherwise, incomplete sealing may cause bleeding." "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating." "when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation." "the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator, forceps, and others).Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/splitting/protruding/partial separating of the tissue pad.In turn, the probe may break before displaying an error window or generating an alarm tone." "do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities." olympus will continue to monitor field performance for this device.
|