• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH DRILL BIT AXSOS 3 TI NON-LOCKING, SHORT, Ø2.5 X 216MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH DRILL BIT AXSOS 3 TI NON-LOCKING, SHORT, Ø2.5 X 216MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 705025
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/15/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "one of our axsos drills has snapped during a case.Drill tip is insitu in tibial plateau and patient has been informed.".
 
Additional Manufacturer Narrative
The reported event could be confirmed based on the inspection performed.The device inspection revealed the following: the identification of the returned drill bit could be confirmed based on the catalog # and lot # marked.The reported event could be confirmed as the instrument is broken at its tip.The breakage surface gives indication of a sudden brittle fracture, resulting most probably from excessive bending and/or torsional load.The area close to the breakage is highly damaged, giving indication that the device was probably already worn out before the reported event.Dimensional and material integrity inspection showed the device to be according to specification.Based on investigation, the root cause was attributed to an user related issue.The failure was most probably caused by excessive bending and/or torsional load.Given the condition of the instrument, it is also probable that hcp failed to properly inspect the drill bit prior to the surgery.As a reminder, the instructions for use clearly state that: "before every operation, ensure that all components and devices to be used during the operation are available in the operation theatre and function correctly with each other.Ensure that drilling and cutting tools are sharp." a review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "one of our axsos drills has snapped during a case.Drill tip is insitu in tibial plateau and patient has been informed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT AXSOS 3 TI NON-LOCKING, SHORT, Ø2.5 X 216MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18884453
Report Number0008031020-2024-00119
Device Sequence Number3051465
Product Code HWE
UDI-Device Identifier07613327083422
UDI-Public07613327083422
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number705025
Device Lot Number25096H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/15/2024
Supplement Date Received by Manufacturer04/10/2024
Initial Report FDA Received Date03/12/2024
Supplement Report FDA Received Date05/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/24/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient SexFemale
-
-