ROCHE DIAGNOSTICS ELECSYS FT3 III; TOTAL TRIIODOTHYRONINE Back to Search Results
Model Number FT3 III
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the customer's e801 analyzer is (b)(6).The ft3 reagent lot number and expiration date used on this analyzer were requested, but not provided.The serial number of the e801 analyzer used for investigation is (b)(6).Ft3 reagent lot number 669112, with an expiration date of 29-feb-2024 was used on this analyzer.The serial number of the e411 analyzer used for investigation is (b)(6).Ft3 reagent lot number 686656, with an expiration date of 30-apr-2024 was used on this analyzer.The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with elecsys ft3 iii on two cobas e 801 module analyzers and a cobas e 411 analyzer.Refer to the attachment for all patient data.The sample was initially tested on the customer's e801 analyzer on (b)(6) 2024.The sample was repeated using the wako accuraseed and abbott alinity methods.The sample was also treated with a 25% polyethylene glycol (peg) solution and repeated on the e801 analyzer.The sample was provided for investigation, where it was tested on a second e801 analyzer and an e411 analyzer on (b)(6)2024.
 
Additional Manufacturer Narrative
Additional investigations of the patient sample determined that it contains an interfering factor against the streptavidin component of the ft3 assay.Per product labeling, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design." the investigation did not identify a product issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT3 III
Common Device Name
TOTAL TRIIODOTHYRONINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18884657
Report Number1823260-2024-00729
Device Sequence Number3089445
Product Code CDP
UDI-Device Identifier07613336171578
UDI-Public07613336171578
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFT3 III
Device Catalogue Number09005811190
Device Lot NumberASKU, 669112, 686656
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/20/2024
Supplement Date Received by Manufacturer06/06/2024
Initial Report FDA Received Date03/12/2024
Supplement Report FDA Received Date06/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age32 YR
Patient SexFemale
-
-