Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.E3: reporter is a j&j employee.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on an unknown date, the staff could not thread the guide wire down back of two viper taps as there was something blocking guide wire from funneling down.The surgeon tried different guidewires (x2) and handles and those were not the issue.The guide wires snapped after trying to get them through tap.Finally the third tap was tried and worked well.There was no surgical delay and no patient consequences/outcome.This report is for one (1) viper2 1.45 dia guidewire, blu this is report 1 of 4 for complaint (b)(4).
|