The investigation of the reported incident was completed by our experts.It was confirmed that the patient suffered a small swelling (2 cm) as a result of the fall.No long-term health consequences were reported.The detailed log-file investigation showed no indication for a system malfunction for the time of the event.In general, tabletops are not designed to prevent patient fall due to required flexibilities for the users.Appropriate material is provided with the system for patient fixation as described in the instructions for use.Additional accessories, e.G.For the treatment of heavy patients and depending on the application of other supporting material, are offered.However, fixation by means of a belt/body straps is essential.The entire process of patient immobilization is the responsibility of the operator.This includes the tabletop, the mattress, the fixation accessories, the alignment of the material and the patient on the table, and the proper execution of the fixation by the operator, including appropriate attention to the patient, e.G., depending on the patient's responsiveness.The concerned artis q ceiling system did not cause or contribute to the patient's fall.
|