PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2 Back to Search Results
Model Number 863283
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A philips field service engineer (fse) went to the customer site and determined the speaker the did not work, and a loudspeaker alarm was triggered.The fse found that it was disassembled, and the speaker connector was loose.The fse reconnected the speaker connector cable.The cause of the reported problem was the connection of the speaker cable.The reported problem was confirmed.No part was replaced.The investigation concludes that no further action is required at this time.Reporting institution phone: (b)(6).Reporter phone: not provided.
 
Event or Problem Description
It was reported the speaker does not work and a loudspeaker alarm is triggered, the device was in clinical use at the time of the event, no adverse event or patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESIGNS VS4 NBP, SPO2
Common Device Name
SURESIGNS VS4 NBP, SPO2
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18886136
Report Number1218950-2024-00154
Device Sequence Number20730697
Product Code DSJ
UDI-Device Identifier00884838028890
UDI-Public00884838028890
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K163649
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number863283
Device Catalogue Number863283
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2024
Initial Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-