All available patient information was included, no additional patient information was available.During the investigation into this issue found that during manufacturing, the microparticle concentrate used in reagent lot number 50808un23 was not properly mixed.As a result, an insufficient amount of microparticles was added to reagent lot 50808un23.A product recall letter was issued on 21feb2024 to all customers who have received shipments of architect stat myoglobin reagent lot numbers 50808un23.The product recall letter notifies the customer of the issue and the potential impact to patient results with use of lot 50808un23.The product recall letter instructs customers to discontinue use of and destroy any remaining inventory of lot 50808un23 according to their laboratory procedures and immediately contact customer support to order a replacement product.
|
|
The customer observed that microparticles in the microparticles bottle appeared lighter than usual, low rlu during calibration and bio rad quality control were out of range low and patient comparison was high while using architect stat myoglobin reagent list (b)(4), lot 50808un23.The following results were provided for sample comparison between 2 reagent lot numbers for myoglobin.50808un23 37319un23 bias sid1: 36.3, 26.6 , -26.72%.Sid2: 117.7, 65.3, -44.52%.Sid3: 256.9, 154.1, -40.02%.Sid4: 364.8, 206.3, -43.45%.Sid5: 810.6, 552, -31.9%.Comparison was completed using patient samples for comparison only and not for the patient management.There was no impact to patient management reported.
|