• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CERAPEDICS INC. I-FACTOR PUTTY; FILLER, BONE VOID, SYNTHETIC PEPTIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CERAPEDICS INC. I-FACTOR PUTTY; FILLER, BONE VOID, SYNTHETIC PEPTIDE Back to Search Results
Model Number 700-050
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Swelling/ Edema (4577)
Date of Event 03/21/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The dhr was reviewed for lot 22c1297 and indicated no abnormal processing.The lot in question met all specifications prior to its release.There have been no ncrs or capas associated with this lot.Cerapedics risk analysis document ra-001, revision 23 (putty) was reviewed.The failure mode identified in this complaint is already identified under line item 38 (potential effect 3) - "inadequate retention in graft site, migration." therefore, no updates to the risk assessment are required.There has been 1 (one) previous complaint related to this potential failure mode.Based on the total number of units distributed (n=(b)(4)), the observed rate is (b)(4) %.The mitigated risk probability for this potential failure mode is estimated to be 1 or an estimated occurrence rate of 0.01%.Therefore, the benefit-risk analysis remains unchanged.Us putty ifu (p/n 40002-07-4) lists the following: "extrusion or migration of the i-factor peptide enhanced bone graft, as is possible with any bone graft, resulting in pain, neural impingement, physical impairment, or loss of function; any of which may require revision surgery".As such, no updates are required for the ifu.
 
Event or Problem Description
Surgeon initially saw the patient on (b)(6) 2023 for a scheduled 2-level acdf at c3/c5.No adverse events occurred during surgery and patient presented with no complications immediately post-op.The patient returned to surgeon's office on (b)(6) 2023, 5 days post-op with significant anterior and left tracheal swelling with tenderness to palpation.The patient's difficulty swallowing raised concern as the patient was unable to eat or drink anything including water.The patient was sent directly back to delray medical center emergency room for admission.Cervical a/p and lateral x-rays show significant pre-vertebral swelling with anterior c2-c7 displacement (27mm) of the trachea without displacement from midline.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-FACTOR PUTTY
Common Device Name
FILLER, BONE VOID, SYNTHETIC PEPTIDE
Manufacturer (Section D)
CERAPEDICS INC.
11025 dover street
suite #1600
westminster CO 80021
Manufacturer Contact
emily aulby
11025 dover street
suite #1600
westminster, CO 80021
MDR Report Key18886502
Report Number3007155473-2024-23131
Device Sequence Number2946586
Product Code NOX
UDI-Device Identifier00850001680028
UDI-Public(01)00850001680028(17)250531(10)22C1297
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P140019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model Number700-050
Device Lot Number22C1297
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2023
Initial Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age58 YR
Patient SexMale
-
-