Blank fields on this form indicate the information is unknown or unavailable.E1: customer name and address = phone: (b)(6), postal code: (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, prior to use for an unknown procedure, a thin strand of plastic came off the side of a cxi support catheter.The catheter was removed from the package, which was not damaged, and was then flushed as normal, without difficulty.Prior to advancement of the catheter over an unspecified wire guide, the catheter was inspected, and a thin strand of plastic was noted to come from the side of the catheter, "about one-third of the total length back from the tip".The catheter was never advanced over the wire.The catheter was not used and never entered the patient.Another catheter was used to complete the procedure.There was no harm to the patient.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Correction: h6 (annex a, annex g).Summary of event: as reported, prior to use for an unknown procedure, a thin strand of plastic came off the side of a cxi support catheter.The catheter was removed from the package, which was not damaged, and was then flushed as normal, without difficulty.Prior to advancement of the catheter over an unspecified wire guide, the catheter was inspected, and a thin strand of plastic was noted to come from the side of the catheter, "about one-third of the total length back from the tip".The catheter was never advanced over the wire.The catheter was not used and never entered the patient.Another catheter was used to complete the procedure.There was no harm to the patient.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions and quality control procedures were conducted during the investigation.A visual inspection of the returned device was conducted as well.The complaint device was returned to cook for investigation.A portion of clear lamination was noted to be stuck on the gold marker band, and skive damage was noted on the catheter.The plastic piece (clear lamination) was able to be removed.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the final lot.A review of complaint history found no additional complaints for the final lot number.Four relevant non-conformances were noted on seven devices from subassembly lots; however, all non-conforming product was scrapped prior to release.No additional relevant complaints were noted on any other final lot associated with the same sub-assembly lots as the complaint device.Cook also investigated all lots manufactured and checked by the same personnel, during the same week, as the complaint device.No additional relevant complaints have been received for any of the lots.The customer followed relevant instructions in the ifu (instructions for use).A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and returned device suggests that there is evidence the device was manufactured out of specification; however, cook has concluded that this is an isolated incident.Although four relevant non-conformances were noted on the sub-assembly lots, all non-conforming product was scrapped, there are 100% inspections in place to capture this non-conformance, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot-related complaints have been received from the field.Therefore, it was concluded that there is no evidence that additional non-conforming product exists in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that manufacturing and quality control deficiencies contributed to this event, as the damage was determined to be caused during the skiving process.Responsible personnel were notified.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|