• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CERAPEDICS INC. I-FACTOR PUTTY; FILLER, BONE VOID, SYNTHETIC PEPTIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CERAPEDICS INC. I-FACTOR PUTTY; FILLER, BONE VOID, SYNTHETIC PEPTIDE Back to Search Results
Model Number 700-050
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Post-operative ct and x-ray images were provided by complainant.Images and relevant details were discussed with an external medical consultant and the following analysis was developed.With regard to causation/relatedness, judgement is that the patient's dysphagia is likely not related to the radiographic findings.Dysphagia is a well characterized complication associated with acdf surgery.The area of compression on the lateral swallow x-ray is at the c5 body, just cephalad the c5-6 disk space.There do not appear to be any significant hyperdensities (i-factor or bone) underlying that area on the 2 sagittal ct images that have been provided, nor does there appear to be any hyperdensity at that area on the swallow x-ray.The compression seen on the lateral swallow x-ray could represent soft-tissue scar, such as "balled up" longus colli muscle overlying the plate or another cause as may be associated with acdf surgery.The radiographic finding under investigation in this case is found at the c7 body, inferior to the plate and just cephalad to the c7-t1 disk space.That is not at the level of compression on the swallow x-ray.The hyperdensities have the appearance of the i-factor particulates/granules, not so much bone/heterotopic bone formation.The "trail" of i-factor to the posterior and then left of the esophagus and trachea is difficult to evaluate on the provided axial cuts, but neither image shows compression of the esophagus or trachea.We do not find connection of this material to the esophagus or vertebral bodies in the images provided to date.Also of note, c7 is not a level typically associated with dysphagia.For all of these reasons, it seems reasonable to conclude the radiographic finding at c7 is unrelated to the patient's dysphagia and posterior esophageal compression seen at c5.With regard to the radiographic finding being unresorbed graft material or ectopic bone, our finding is that this is unresorbed graft material.This finding is consistent with long standing observations in the worldwide clinical history of i-factor since 2008 as well as ectopic bone formation specific preclinical studies.These images collectively and individually don't suggest the presence of ectopic bone formation and/or densification and consolidation.There is no evidence of capsule, neocortex, inflammatory changes surrounding the tail etc.The material in the axial image at the second time point looks consistent with material not connected to any boney structure and which is comprised of distinct radiodense particles and clumps of particles.This is consistent with i-factor clinical experiences as well as animal models demonstrating i-factor purposefully in soft tissue does not elicit ectopic bone formation."superior laryngeal nerve injury and dysphagia" are listed as potential adverse events in the ifu.Additional information: most recent ct scan on (b)(6) 2023 - "unchanged linear hyperdensity material extending from the posterior aspect of the proximal esophagus at the level of c7-t1 and tracking along the left tracheoesophageal groove medial to the left carotid space.Findings again probably postsurgical in nature.".
 
Event or Problem Description
Patient presented with migration of the bone void filler from the graft site, resulting in patient pain and dysphagia.Patient contacted cerapedics after closure of c22-110 complaining of on-going neck pain near c7 on the left lower side stating that they have gotten 4 ct scans in less than 28 mos and several other tests.The ct results remain unchanged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-FACTOR PUTTY
Common Device Name
FILLER, BONE VOID, SYNTHETIC PEPTIDE
Manufacturer (Section D)
CERAPEDICS INC.
11025 dover street
suite #1600
westminster CO 80021
Manufacturer Contact
emily aulby
11025 dover street
suite #1600
westminster, CO 80021
MDR Report Key18886676
Report Number3007155473-2024-23133
Device Sequence Number7774074
Product Code NOX
UDI-Device Identifier00850001680028
UDI-Public(01)00850001680028(17)240531(10)21C0969
Combination Product (Y/N)Y
PMA/510(K) Number
P140019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model Number700-050
Device Lot Number21C0969
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2023
Initial Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN CERVICAL PLATE SYSTEM
Outcome Attributed to Adverse Event Other;
Patient Age66 YR
Patient SexFemale
-
-