• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CHECK VALVE WITH MALE/FEMALE LUER LOCK; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. CHECK VALVE WITH MALE/FEMALE LUER LOCK; STOPCOCK, I.V. SET Back to Search Results
Catalog Number 011-CS25
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
The event involved a check valve with male/female luer lock where it was reported by the customer the check valve was nonfunctional.The infused product does not pass, the syringe must be pushed hard.Verification of the catheter on the patient, which is functional.The patient could have been reinfused if the reporter had not tested the catheter.The valve was removed and discarded.The incident happened at the interventional preparation room.The status of the product at the time of the event is during infusion.There was patient involvement and no report of patient harm.
 
Additional Manufacturer Narrative
The complaint of no flow / can't prime / difficult to prime on item 011-cs25 cannot be confirmed by investigation.Since no product samples, pictures, or videos were received for investigation.Without the return of the used sample a comprehensive failure investigation cannot be performed, and a probable cause cannot be determined.The device history review (dhr) lot#13752166 was reviewed and no nonconformities were found that would have led the reported condition on the complaint.E1initial reporter address: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHECK VALVE WITH MALE/FEMALE LUER LOCK
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18886719
Report Number9617594-2024-00256
Device Sequence Number2997820
Product Code FMG
UDI-Device Identifier00840619006075
UDI-Public(01)00840619006075(17)280901(10)13752166
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K964435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number011-CS25
Device Lot Number13752166
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2024
Initial Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYRINGE, UNK MFR; UNSPECIFIED CATHETER, UNK MFR
Patient SexUnknown
-
-