• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFEN XTAB 6X45MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER PRIME CFXFEN XTAB 6X45MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 186760445
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Ambulation or Postural Difficulties (2544); Paresthesia (4421)
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in brazil as follows: it was reported that a procedure of lumbar arthrodesis was performed on a patient.After 10 months of the procedure, the patient came into contact with the doctor complaining of strong lumbar pain, with irradiation to the left lower limb and paresthesia associated with important functional limitation, with difficulty performing daily work activities.Clinical and radiological tests were performed and proved the screw breakage.A new procedure was carried out for removing the screw and implanting a new one.This report involves one (1) viper prime cfx fen x-tab polyaxial screw 5.5 6 x 45mm.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for viper prime cfxfen xtab 6x45mm, was conducted identifying that lot number tbaish was released in three batches.¿ batch1: lot qty of (b)(4) units were released on 28 sept 2021 with no discrepancies.¿ batch2: lot qty of (b)(4) units were released on 01 nov 2021 with no discrepancies.¿ batch3: lot qty of (b)(4) units were released on 07 mar 2022 with no discrepancies.Supplier: depuy: tomz corp.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that one of the two screws shown on image is broken from the threaded upper middle part.Broken fragment is not observed.With the available information, it is not possible to establish a potential cause for the failure of the device.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the viper prime cfxfen xtab 6x45mm would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d9 h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that the screw was broken from the threaded upper middle part.Broken fragment was not returned for analysis.Furthermore, evidence of normal wear consistent with implantation and explantation during surgical technique was observed.As well, a foreign substance was also found, potentially being a biological residue.While no potential cause can be established with the available information, factors such as the age of the patient, underlying disease, bone density, or a possible early weight-bearing, could have led to failure of the implant.A dimensional inspection was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the viper prime cfxfen xtab 6x45mm would contribute to the complained device issue.Based on the investigation findings, the potential cause cannot be established, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME CFXFEN XTAB 6X45MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
chemin blanc 38
le locle 02400
SZ   02400
3035526892
MDR Report Key18887271
Report Number1526439-2024-01029
Device Sequence Number11531725
Product Code NKB
UDI-Device Identifier10705034509402
UDI-Public10705034509402
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K170543
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number186760445
Device Lot NumberTBAISH
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/20/2024
Supplement Date Received by Manufacturer03/14/2024
06/12/2024
Initial Report FDA Received Date03/12/2024
Supplement Report FDA Received Date03/29/2024
07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-