| Catalog Number |
03-2622-3 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
02/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility physician reported to fresenius that a broken line was detected on the combiset bloodlines during the patient's hemodialysis (hd) treatment.There was no serious injury or required medical intervention reported.The patient's estimated blood loss (ebl) was not provided.No parts were available to be returned to the manufacturer.Additional information was requested however a response was not received.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A user facility physician reported to fresenius that a broken line was detected on the combiset bloodlines during the patient's hemodialysis (hd) treatment.There was no serious injury or required medical intervention reported.The patient's estimated blood loss (ebl) was not provided.No parts were available to be returned to the manufacturer.Additional information was requested however a response was not received.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer for physical evaluation.However a photograph was provided by the customer depicting the saline "t" connector detached from the combi set.The reported event was confirmed in accordance with the picture received.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.
|
| |
|