| Model Number |
N/A |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).G2 foreign: japan.Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during incoming inspection the product was found to be nonconforming.The product had a foreign substance within the packaging.There was no patient involvement.Diligence is complete.No additional information is available.No adverse events were reported as a result of this malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded by zimmer biomet under (b)(4).Visual inspection confirmed there was debris sealed inside the package of the connector.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional information is available regarding the incident.
|
| |
|