| Model Number |
MAF-GM |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned, and the evaluation found no malfunctions aside from the allegation reported in b5.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the airway mobilscope exhibited peeling of the tube coating.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.It is likely the reported event was caused by stress from repeated use, external factors, or handling.The event can be detected by handling the device in accordance with the following instructions for use (ifu): chapter 3 preparation and inspection, 3.6 inspection of the endoscope olympus will continue to monitor field performance for this device.
|
| |
|