AIZU OLYMPUS CO., LTD. EVIS EXERA III COLONOVIDEOSCOPE Back to Search Results
Model Number PCF-PH190I
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during reprocessing, the colonovideoscope tested positive for >80 colony forming units (cfus) of serratia marcescens.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Additional Manufacturer Narrative
The suspect device has been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.The legal manufacture reviewed the customer provided the cleaning disinfection and sterilization (cds) processes where [no obvious deviations from instructions for use (ifu) were identified.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on results of third-party testing, the customer provided cleaning disinfection and sanitization (cds) practices, the device evaluation, and the legal manufacturer's final investigation.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: 1cfu.Bacterial identification: serratia marcescens.Sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: 1cfu.Bacterial identification: bacillaceae.It is unclear if reprocessing was carried out according to the instructions for use.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu) before repair, growth was observed but the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18891362
Report Number9610595-2024-05294
Device Sequence Number15350613
Product Code FDF
UDI-Device Identifier04953170436710
UDI-Public04953170436710
Combination Product (Y/N)N
Initial Reporter CountryLU
PMA/510(K) Number
K131780
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPCF-PH190I
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date03/12/2024
Supplement Report FDA Received Date05/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-