It was reported that after an initial bunion surgery, all hardware was removed on (b)(6) 2024 due to pain and irritation.One screw broke while being removed, however additional information confirms the broken screw was a result of removal efforts.There was no report of a deficiency alleged/found against any of the tmc devices prior to removal of the hardware.No other patient outcomes were reported as a result of this event and the patient's bones were completely fused/healed.No devices were returned for evaluation.Device-specific information was not available; therefore, a review of device history records was not able to be performed.However, all non-conformances for possible kits utilized in surgery were reviewed and no non-conformances or issues during the manufacture or release of the products were identified to date that could have contributed to what was reported.Although the most likely cause of the reported event could not be determined with certainty without return of the device, it was additionally reported that rubbing of the dorsal plate may have contributed to what the patient experienced.There were no deficiencies or malfunctions alleged/found with any tmc device, nor were there any reported issues with placement of the device or healing of the surgical site.The company will supplement this mdr as necessary and appropriate.
|