| Model Number |
FINETUNER ECHO [FINETUNER 2] |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
A finetuner echo was returned to service and repair.Upon preliminary investigation at service and repair it was determined that there is failure of the tantalum capacitor.According to the repair request form, the device had no function.No injury has been reported.
|
| |
|
Event or Problem Description
|
|
A finetuner echo was returned to service and repair.Upon preliminary investigation at service and repair it was determined that there is failure of the tantalum capacitor.According to the repair request form, the device was mechanically not functioning.No injury has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according the information received, a finetuner echo was sent to service _ repair due to mechanical problems.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.No injury was reported.This is a final report.
|
| |
|
Search Alerts/Recalls
|