Catalog Number 8607300
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the machine rebooted five times during the case and each time booted up to man/spont mode.No injury reported.
 
Additional Manufacturer Narrative
The device log file was requested but was not provided for investigation.Therefore, a case specific analysis was not possible.The device was tested and inspected on site.All tests passed and the reported symptom could not be reproduced.Finally, based on the available information the root cause for the reported symptom could not be determined.A restart as reported is obvious for the user, since the device is intended to be used under constant supervision only.During a restart, automatic ventilation is interrupted shortly.Manual ventilation via the integrated manual breathing bag remains possible at any time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS MRI
Common Device Name
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key18892676
Report Number9611500-2024-00108
Device Sequence Number17745043
Product Code BSZ
UDI-Device Identifier04048675041498
UDI-Public(01)04048675041498(11)201119(17)210316(93)8607300-48
Combination Product (Y/N)N
PMA/510(K) Number
K072884
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8607300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/14/2024
Initial Report FDA Received Date03/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/19/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-