STRYKER ORTHOPAEDICS-MAHWAH X3 TRIATHLON INSERT CS NUM6 10MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5531-G-610-E |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
As reported: "on (b)(6) 2024 we were scheduled for a retinacular repair and i&d of a right total knee arthroplasty.Upon exposure to the joint, dr.Noticed that the 6x10 tibia insert was "rocking" in the tibial tray.After inspection of the insert, it appeared as if it was correctly engaged to the tibial baseplate, however was still rocking posteriorly.We took note of this and assumed that soft tissue was preventing the tibial insert from fully locking into the baseplate.We proceeded forward with the explant of the device, upon which dr.Examined the explanted 6x10 insert.Here, he noticed two "bevels on the posterior aspects of the tibial insert.He expressed his concerns that he thought something wasn't correct.He continued with his soft tissue work, i&d, and we proceeded to trial for a new poly.Once we chose our new cs poly, he compared the two on the back table.It was here that we noticed a difference in the two implants.The new 6x12 cs insert did not have the "bevels" posteriorly like the explanted 6x10 cs did.We thought maybe it was damaged when initially implanted but the bevels were oddly symmetric on both the medial and lateral aspect of the implant.Upon implantation of the new 6x12 cs insert, the poly was stable and the proceeded to close.He believes the unstable 6x10 poly caused the patient pain/instability)." rep confirmed only the insert was revised.Surgeon would like to know if this issue has been reported previously.
|
| |
|
Event or Problem Description
|
|
As reported: "on 2/21/2024 we were scheduled for a retinacular repair and i&d of a right total knee arthroplasty.Upon exposure to the joint, dr.Noticed that the 6x10 tibia insert was "rocking" in the tibial tray.After inspection of the insert, it appeared as if it was correctly engaged to the tibial baseplate, however was still rocking posteriorly.We took note of this and assumed that soft tissue was preventing the tibial insert from fully locking into the baseplate.We proceeded forward with the explant of the device, upon which dr.Examined the explanted 6x10 insert.Here, he noticed two "bevels on the posterior aspects of the tibial insert.He expressed his concerns that he thought something wasn't correct.He continued with his soft tissue work, i&d, and we proceeded to trial for a new poly.Once we chose our new cs poly, he compared the two on the back table.It was here that we noticed a difference in the two implants.The new 6x12 cs insert did not have the "bevels" posteriorly like the explanted 6x10 cs did.We thought maybe it was damaged when initially implanted but the bevels were oddly symmetric on both the medial and lateral aspect of the implant.Upon implantation of the new 6x12 cs insert, the poly was stable and the proceeded to close.He believes the unstable 6x10 poly caused the patient pain/instability)." rep confirmed only the insert was revised.Surgeon would like to know if this issue has been reported previously.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding disassociation/damage involving a triathlon insert was reported.The event was confirmed via evaluation of the returned device.Method & results: product evaluation and results: visual inspection: visual inspection of the returned device indicated that the device is damaged in the area of the locking mechanism, consistent with disassembly through explantation.The locking wire was returned separated from the device.The posterior edges of the device do exhibit similar appearing "bevels" consistent with that reported in the event description.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to instability and pain.Intraoperatively, the 6x10 tibia insert was observed to be "rocking" in the tibial tray.Visual inspection of the returned device indicated that the device is damaged in the area of the locking mechanism, consistent with disassembly through explantation.The locking wire was returned separated from the device.The posterior edges of the device do exhibit similar appearing "bevels" consistent with that reported in the event description.The triathlon knee system surgical protocol, document # (b)(4), indicates that to properly assemble the tibial insert to the baseplate, the surgeon must "distract the joint and angle the insert posteriorly into the primary tibial baseplate.The posterior lip of the tibial insert must fit beneath the lip on the posterior primary tibial baseplate wall." the insert must then be impacted in place anteriorly using the tibial insert impactor (attached to the impaction handle).The appropriate tibial insert is fully seated once the locking wire locks under the barbs on the anterior/interior surface of the tibial baseplate wall.The damage present on the complaint device appears consistent with the surgeon angling the anterior side of the tibial insert into the baseplate first, instead of the posterior side as directed in the surgical protocol.The observed damage is likely related to this surgical technique during assembly.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|