| Catalog Number |
365978 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd microtainer® sst¿- amber, poor barrier separation occurred with one tube.The tube was recentrifuged and the barrier still did not move, causing the serum to not be able to be analyzed.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, fifty(50) retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to poor barrier separation as all samples met specifications.Complaints for sample quality are under statistical control for the month of february 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode poor barrier separation.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Search Alerts/Recalls
|