• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE PS FEM RT SZ 5 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE PS FEM RT SZ 5 NAR CEM; ATTUNE IMPLANT : KNEE FEMORAL Back to Search Results
Catalog Number 150410225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Synovitis (2094); Discomfort (2330); Joint Laxity (4526)
Date of Event 03/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2015, the patient underwent a right total knee arthroplasty utilizing depuy synthes implants including patella and cement x 2.On (b)(6) 2023, the patient underwent a right knee revision secondary to discomfort, falls, joint instability, and a positive bone scan with suspected tibial loosening.Intraoperatively, hypertrophic synovial and granulation tissue was identified.Loosening at the implant to cement interface was confirmed.(b)(6) 2015 doi.(b)(6) 2023 dor (right knee).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM RT SZ 5 NAR CEM
Common Device Name
ATTUNE IMPLANT : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18894585
Report Number1818910-2024-05708
Device Sequence Number6130537
Product Code JWH
UDI-Device Identifier10603295041849
UDI-Public10603295041849
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue Number150410225
Device Lot Number7997646
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/06/2024
Supplement Date Received by Manufacturer04/11/2024
Initial Report FDA Received Date03/13/2024
Supplement Report FDA Received Date04/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/03/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE FB TIB BASE SZ 4 CEM; ATTUNE MEDIAL DOME PAT 35MM; ATTUNE PS FB INSRT SZ 5 5MM; SMARTSET GMV 40G US EO; SMARTSET GMV 40G US EO
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-