• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES C4121, 45CM GRASPER REPOS CART, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES C4121, 45CM GRASPER REPOS CART, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC Back to Search Results
Model Number C4121
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Type of procedure: ni.Event description: hospital: [hospital].Product: 45cm grasper."robotic case.Blue pads (reporter assumes both blue pads of the grasper) detached and fell into the patient.The surgeon was able to retrieve the pads." products available for return.Patient status: no patient injury or illness occurred associated with the complaint event.Intervention: ni.
 
Additional Manufacturer Narrative
The event device has returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of pad detachment.Based on the condition of the returned unit, it is likely that the pad detachment occurred due to adhesive failure.It is also likely that this occurred during usage.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Type of procedure: ni.Event description: hospital: [hospital].Product: 45cm grasper."robotic case.Blue pads (reporter assumes both blue pads of the grasper) detached and fell into the patient.The surgeon was able to retrieve the pads." products available for return.Patient status: no patient injury or illness occurred associated with the complaint event.Intervention: ni.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C4121, 45CM GRASPER REPOS CART, 10/BX
Common Device Name
LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key18895138
Report Number2027111-2024-00435
Device Sequence Number12324306
Product Code NWV
UDI-Device Identifier00607915116873
UDI-Public(01)00607915116873(17)260725(30)01(10)1496447
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2026
Device Model NumberC4121
Device Catalogue Number101472473
Device Lot Number1496447
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/13/2024
Supplement Report FDA Received Date04/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-