| Model Number |
C4121 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Type of procedure: ni.Event description: hospital: [hospital].Product: 45cm grasper."robotic case.Blue pads (reporter assumes both blue pads of the grasper) detached and fell into the patient.The surgeon was able to retrieve the pads." products available for return.Patient status: no patient injury or illness occurred associated with the complaint event.Intervention: ni.
|
| |
|
Additional Manufacturer Narrative
|
|
The event device has returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of pad detachment.Based on the condition of the returned unit, it is likely that the pad detachment occurred due to adhesive failure.It is also likely that this occurred during usage.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|
Event or Problem Description
|
|
Type of procedure: ni.Event description: hospital: [hospital].Product: 45cm grasper."robotic case.Blue pads (reporter assumes both blue pads of the grasper) detached and fell into the patient.The surgeon was able to retrieve the pads." products available for return.Patient status: no patient injury or illness occurred associated with the complaint event.Intervention: ni.
|
| |
|