| Catalog Number |
367922 |
| Medical Device Problem Code |
Coagulation in Device or Device Ingredient (1096)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® sodium fluoride potassium oxalate blood collection tubes, the samples were clotted.There was no report of impact to the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields have been updated with additional/corrected information: d.9 device available for eval: yes.D.9 returned to manufacturer on: 29-feb-2024.H.6.Investigation summary: bd received 5 samples for investigation.The samples and retention samples were visually inspected with no issues observed.Complaints for sample quality are under statistical control for the month of february 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode , clotting.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® sodium fluoride potassium oxalate blood collection tubes, the samples were clotted.There was no report of impact to the patient or user.
|
| |
|