| Model Number |
CLH-SC |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the halogen light source exhibited the ceramic lamp socket unit was loose when connected to halogen lamp.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and the evaluation found no additional reportable malfunctions other than those mentioned in b5.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to a defective lamp socket.Olympus will continue to monitor field performance for this device.
|
| |
|