|
DEPUY SPINE INC PWR XPDM POLY DRIVER, HANDLE; INSTR, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR/ACCESS & ATTACH
|
Back to Search Results |
|
| Catalog Number |
310002150P |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on february 26 2024, the screwdriver distal end tip was stripped and the sleeve no longer engaged with the screw head correctly.The patient was not impacted as a new screwdriver from another set was used.There was no delay in surgery.This report is for one pwr xpdm poly driver, handle this is report 1 of 3 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation did not revealed the reported stripped and unable to assemble condition for pwr xpdm poly driver, handle however it shows the signs of usage.Review of provided photo shows the device however it does not confirms the reported stripped and unable assemble condition and cause remains undetermined.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the pwr xpdm poly driver, handle would not contribute to the complained device issue.Based on the investigation findings, potential cause not established and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|