| Model Number |
7590 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2024.It was reported that crossing difficulties and shaft kink occurred.The severely calcified target lesion was located in the left anterior descending artery.A 2.00mm x 20mm maverick balloon catheter was advanced but failed to cross the lesion, and the shaft catheter was kinked.The procedure was completed with another of the same device.No patient complications were reported, and the patient condition was stable.However, returned device analysis revealed balloon longitudinal tear.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a maverick 2 mr balloon catheter.A visual inspection revealed no damage or irregularity.A microscopic inspection revealed a 9mm longitudinal tear at the distal waist.There was blood present in the guidewire lumen, as well as blood and contrast present in the inflation lumen and the loosely folded balloon.
|
| |
|