| Catalog Number |
72205610 |
| Medical Device Problem Codes |
Break (1069); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the disposable blade that was inserted into the quad cutter became stuck and could not be removed.The disposable blade was bent and did not come out despite multiple attempts.No case reported; therefore, there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The outer blade, proximal cutter is fractured away from the shaft with the proximal portion of a disposable blade in the channel.The metal of the proximal portion of the blade is mangled and the device is scratched and gouged.A functional evaluation could not be performed due to the condition in which the device was received.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, attempted correction of a damaged device, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|