| Catalog Number |
72202897 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unspecified arthroscopy, the twinfix ultra ti anchor did not hold on to the bone and it came out of the cortex.The procedure was completed with a s+n back up device.There was a delay less than 30 minutes and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|