• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39575
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: initial reporter address 1:(b)(6).E1: initial reporter city: (b)(6).Device evaluated by mfr.: the amplatz super stiff guidewire was returned for analysis.A visual examination observed that the guidewire was bent and stretched at the distal tip section.Tip detachment was not confirmed.
 
Event or Problem Description
It was reported that tip detachment occurred.A 035/260/1 (bx/1) amplatz super stiff guidewire was selected for use.The device was unpacked and attempted to be inserted into the sheath, the tip was noted to be chipped.A second guidewire was used to complete the procedure.No complications were reported.
 
Event or Problem Description
It was reported that tip detachment occurred.A 035/260/1 (bx/1) amplatz super stiff guidewire was selected for use.The device was unpacked and attempted to be inserted into the sheath, the tip was noted to be chipped.A second guidewire was used to complete the procedure.No complications were reported.
 
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).Initial reporter city: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ SUPER STIFF
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18899128
Report Number2124215-2024-14255
Device Sequence Number7683029
Product Code DQX
UDI-Device Identifier08714729295655
UDI-Public08714729295655
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K843012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number39575
Device Catalogue Number39575
Device Lot Number0032712373
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/19/2024
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date03/13/2024
Supplement Report FDA Received Date04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-