| Model Number |
39575 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter address 1:(b)(6).E1: initial reporter city: (b)(6).Device evaluated by mfr.: the amplatz super stiff guidewire was returned for analysis.A visual examination observed that the guidewire was bent and stretched at the distal tip section.Tip detachment was not confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that tip detachment occurred.A 035/260/1 (bx/1) amplatz super stiff guidewire was selected for use.The device was unpacked and attempted to be inserted into the sheath, the tip was noted to be chipped.A second guidewire was used to complete the procedure.No complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that tip detachment occurred.A 035/260/1 (bx/1) amplatz super stiff guidewire was selected for use.The device was unpacked and attempted to be inserted into the sheath, the tip was noted to be chipped.A second guidewire was used to complete the procedure.No complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address 1: (b)(6).Initial reporter city: (b)(6).
|
| |
|