• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO XL; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO XL; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00547110
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a040609 captures the reportable event of balloon tip bent.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro xl retrieval balloon was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During the procedure, the patient was examined using an extractor balloon.Although the wire was already in the bile duct, the extractor balloon could not be inserted over the wire.An attempt was made to pull the wire back a little, but unfortunately, it proved to be a bit difficult.After removing the extractor balloon, it was discovered that the distal tip of the balloon was bent.The procedure was completed with another extractor pro xl retrieval balloon.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR PRO XL
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18899549
Report Number3005099803-2024-00985
Device Sequence Number18404441
Product Code FGE
UDI-Device Identifier08714729790402
UDI-Public08714729790402
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K102082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/12/2025
Device Model NumberM00547110
Device Catalogue Number4711
Device Lot Number0033053598
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/22/2024
Initial Report FDA Received Date03/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-