MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT CATHETER; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HEMOSPLIT CATHETER; DIALYSIS CATHETER Back to Search Results
Catalog Number 5733690
Medical Device Problem Codes Material Discolored (1170); Material Opacification (1426); Stretched (1601); Material Protrusion/Extrusion (2979)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 01/2025).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that six months and one day post a dialysis catheter placement, the catheter hubs were allegedly getting weaker, causing ballooning while flushing and also catheter hub color was allegedly changed to pale yellow.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were provided for review.The photo shows catheter discoloration on both luer.Catheter bulging was noted to the blue luer near the clamp.Therefore, the investigation is confirmed for the reported material discoloration, material stretch and protrusion.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 01/2025), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that six months and one day post a dialysis catheter placement, the catheter hubs were allegedly getting weaker, causing ballooning while flushing.It was further reported that the color of the catheter hub was allegedly changed to pale yellow.There was no reported patient injury.
 
Event or Problem Description
It was reported that six months and one day post a dialysis catheter placement, the catheter hubs were allegedly getting weaker, causing ballooning while flushing.It was further reported that the color of the catheter hub was allegedly changed to pale yellow.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: although the physical device was not returned, two electronic photos were provided for review.The first photo shows catheter discoloration on both luer.Catheter bulging was noted to the blue luer near the clamp.Both extension legs were opaque near the bifurcation area.The second photo shows the xepi skin prep antiseptic liquid bottle that was not included in the prohibited chemicals to be used on the catheter as per ifu, but the fluids used are supposed to be left to dry before the device is dressed again.Therefore, the investigation is confirmed for the reported material discoloration, material stretch, protrusion issues and identified material opacification.A definitive root cause could not be determined based upon the provided information.Labeling review: adequate a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Ifu reviewed: baw0727959, rev.0 ifu warns: *acetone and peg-containing ointments can cause failure of this device and should not be used with polyurethane catheters.Chlorhexidine patches or bacitracin zinc ointments (e.G., polysporin* ointment) are the preferred alternative.*alcohol or alcohol-containing antiseptics (such as chlorhexidine) may be used to clean the catheter/skin site; however, care should be taken to avoid prolonged or excessive contact with the solutions(s).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPLIT CATHETER
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18899908
Report Number3006260740-2024-01073
Device Sequence Number15413833
Product Code MSD
UDI-Device Identifier00801741013096
UDI-Public(01)00801741013096
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K030020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Catalogue Number5733690
Device Lot NumberREHQ4676
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/16/2024
Supplement Date Received by Manufacturer04/22/2024
08/23/2024
Initial Report FDA Received Date03/13/2024
Supplement Report FDA Received Date04/25/2024
09/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-