H3: the axium prime coil was returned for analysis.The excelsior sl-10 micro catheter used in the event was not returned.The excelsior sl-10 micro catheter has a labeled id (inner diameter) of 0.0165¿, as per an online source.Per the axium prime coil ifu (instructions for use): ¿axium prime detachable coils should be delivered through a microcatheter with a minimum inside diameter of 0.0165¿-0.017¿ with two marker bands¿ therefore, the excelsior sl-10 micro catheter was found to be compatible for use with the axium prime coil.The axium prime implant coil was returned within its introducer sheath.The axium prime coil was found to be inverted within the introducer sheath with the wavelock at the distal end.The pushwire was found to be bent at 68.0cm from proximal end.The axium prime implant coil was found to be broken/missing.However, the polypropylene still intact.No damages or irregularities were found with the introducer sheath.No other anomalies were observed.The axium prime coil was pushed out from introducer sheath without any issues.The axium prime coil could not be used for testing with an in-house microcatheter due to its damaged condition.Based on the device analysis and the reported information, the customer¿s report of ¿coil resistance /stuck in catheter hub¿ was con firmed as the axium prime coil was found to be damaged.However, the root cause could not be determined as the axium prime coil could not be used for resistance testing with an in-house micro catheter due to its damaged condition.Possible causes for coil resistance at catheter hub are, but not limited to, failure to hydrate the coil prior to use, failure to utilize continuous flush, failure to use a compatible microcatheter for delivery, and use of an improper hubbing technique (over tightening of the rhv/sheath firmly seated into hub).Improper hubbing technique could not be ruled out as a potential root cause for the event.The customer reported preparing the axium prime coil prior to use (which includes implant coil hydration).The excelsior sl-10 catheter was found to be compatible for use with the axium prime coil.Information regarding if a continuous flush was used was not reported.Therefore, any contributing factors could not be assessed.It is possible that the damage to the implant coil and the pushwire occurred during the attempt to push the coil through the micro catheter against the reported resistance.However, the cause of resistance could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|