| Model Number |
X SERIES |
| Medical Device Problem Code |
Accessory Incompatible (1004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was verified and attributed to a damaged defib receptacle.The defib receptacle was replaced to remedy the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device was unable to connect to the multifunction cable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|