| Model Number |
CCWTT3 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Impairment (2138); Visual Disturbances (2140)
|
| Date of Event |
04/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced blurry vision, halos, dull color vision and was not able to drive at night, dysphotopsia.The lens was exchanged for another lens model 11 months 21 days following the initial procedure.There are two medical device reports associated with this patient.This report is associated with the right eye.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned for evaluation.Solution is dried on the lens.The lens is cut in half, typical of insertion and removal from the eye.Marks are observed on both sides of the optic that are similar to those left by a surgical instrument used to grasp the lens.A power and resolution inspection could not be conducted due to extensive damage.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.The product investigation could not identify a root cause for the reported complaint.A power and resolution inspection could not be conducted due to extensive damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.Information was provided that the company model (1.50cyl) 22.5 diopter (add power +2.2/+3.2) lens was replaced with a company model 22.0 diopter lens.Due to the exchange of a toric multifocal lens to a non-toric monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.The manufacturer internal reference number is: (b)(4).
|
| |
|