Catalog Number 198-37L
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the white cuff inflated without any problem, but the white cuff pilot balloon would not inflate.There was patient involvement but no patient harm or adverse event was reported.
 
Additional Manufacturer Narrative
B3: month and year of event have been provided, day is unknown.A product sample was received and is awaiting evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
H3 and h6.Evaluation codes: updated.One sample was returned without the original package.During visual inspection no discrepancies were detected.Per functional testing, air was introduced through the blue and white pilot balloon to inflate them.The cuffs inflated successfully; the complaint was not confirmed/duplicated.Root cause cannot be addressed since the complaint was not confirmed.No action taken.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX ENDOBRONCHIAL TUBE, 37 FRENCH
Common Device Name
TUBE, TRACHEAL/BRONCHIAL, DIFFERENTIAL VENTILATION
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18902497
Report Number9617604-2024-00230
Device Sequence Number21094756
Product Code CBI
UDI-Device Identifier15019315058622
UDI-Public(01)15019315058622(17)251008(10)4325778
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K953483
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/08/2025
Device Catalogue Number198-37L
Device Lot Number4325778
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/26/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/15/2024
Supplement Date Received by Manufacturer04/17/2024
Initial Report FDA Received Date03/14/2024
Supplement Report FDA Received Date05/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-