| Catalog Number |
198-37L |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the white cuff inflated without any problem, but the white cuff pilot balloon would not inflate.There was patient involvement but no patient harm or adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: month and year of event have been provided, day is unknown.A product sample was received and is awaiting evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 and h6.Evaluation codes: updated.One sample was returned without the original package.During visual inspection no discrepancies were detected.Per functional testing, air was introduced through the blue and white pilot balloon to inflate them.The cuffs inflated successfully; the complaint was not confirmed/duplicated.Root cause cannot be addressed since the complaint was not confirmed.No action taken.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.
|
| |
|