| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
In march 2024, the recipient stopped hearing with the device after having bumped her head.Re-implantation has been scheduled for (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The active electrode also showed some additional damage to the electrode wires, as being caused by minute device mobility.An accident or impact might have weakened the fixation of the implant, consequently leading to electrode mobility.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
In (b)(6) 2024, the recipient stopped hearing with the device after having bumped her head.The user has been re-implanted.
|
| |
|