• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60WF
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 09/08/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, lens was explanted at secondary procedure due to patient experienced decreased quality of vision, blurry vision even with glasses.The iol was replaced with other company lens approximately one year after initial procedure.There are two medical device reports associated with this patient.This report is associated with the left eye.
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution is dried on the lens.The lens is cut in half, typical of insertion and removal from the eye.A deep scratch is observed on the optic near the edge of the lens.A power and resolution inspection could not be conducted due to extensive damage.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.The product investigation could not identify a root cause for the reported complaint.A power and resolution inspection could not be conducted due to extensive damage.The power and resolution of each lens is 100% evaluated during the manufacturing process to determine acceptability per model and diopter.Information was provided that the company 24.5 diopter lens was replaced with a non-company monofocal 24.0 diopter lens.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF ASPHERIC UV ABSORBING SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18902806
Report Number1119421-2024-00483
Device Sequence Number9068984
Product Code HQL
UDI-Device Identifier00380652332750
UDI-Public00380652332750
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberSA60WF
Device Catalogue NumberSA60WF.245
Device Lot Number12671789
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/16/2024
Supplement Date Received by Manufacturer04/05/2024
Initial Report FDA Received Date03/14/2024
Supplement Report FDA Received Date05/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LAL 24.0 70045420614.; MONARCH II LOADING FORCEPS.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-